First adjustment can be done 4 to 8 weeks
post-op (1-2 ml) According to weight evolution 0.4 ml = 0.5 mm adjusted Always use Huber tip Use sterile saline Check volume before adjustments Use a 3-5 ml Syringe Maximum fill is 4 cc
X-ray of abdomen Barium swallow Patient standing up Check passage and pouch size Patient in decubitus, place Access Port needle
under flouroscopic control Check volume inside the band before adjusting